Emerson 1700 Satellite Radio User Manual


 
Installation Manual 43
Output Wiring – Model 2700 Configurable I/O Transmitters
Output Wiring – Configurable I/O SpecificationsOutput Wiring – Fieldbus/PROFIBUSOutput Wiring – Intrinsically Safe
Figure 6-2 Basic mA wiring
Figure 6-3 HART/analog single-loop wiring
+
mA1
mA2
+
820 Ω maximum loop resistance
420 Ω maximum loop resistance
820 Ω maximum loop resistance
For HART communications:
•600 Ω maximum loop resistance
•250 Ω minimum loop resistance
+
HART-
compatible host
or controller