Emerson 1700 Satellite Radio User Manual


 
44 Micro Motion
®
Model 1700 and 2700 Transmitters
Output Wiring – Model 2700 Configurable I/O Transmitters
Figure 6-4 HART multidrop wiring with SMART FAMILY
transmitters and a configuration tool
6.4 Frequency output wiring
Frequency output wiring depends on whether you are wiring terminals 3 and 4 (Channel B) or
terminals 5 and 6 (Channel C), and also on whether you have configured the terminals for internal or
external power. The following diagrams are examples of proper wiring for these configurations:
Channel B, internal power – Figure 6-5
Channel B, external power – Figure 6-6
Channel C, internal power – Figure 6-7
Channel C, external power – Figure 6-8
Note: If both Channel B and Channel C are configured for frequency output, the Channel C signal is
generated from the Channel B signal, with a user-specified phase shift. The signals are electrically
isolated but not independent. This configuration is used to support dual-pulse and quadrature modes.
For more information, see the manual entitled Series 1000 and 2000 Transmitters Configuration and
Use Manual.
HART-compatible
host or controller
HART-compatible
transmitters
SMART FAMILY
transmitters
Note: For optimum HART communication,
make sure the output loop is
single-point-grounded to an
instrument-grade ground.
24 VDC loop power
supply required for
HART 4–20 mA
passive transmitters
Model 2700
configurable I/O
transmitter
(internally powered
outputs)
600 Ω maximum resistance
250 Ω minimum resistance