Step 1: Confirm system readiness
Confirm HART revision capability
If using HART-based control or asset management systems, confirm the HART
capability of those systems prior to transmitter installation. Not all systems
are capable of communicating with the HART Revision 7 protocol.
This transmitter can be configured for either HART Revision 5 or 7.
For instructions on how to change the HART revision of your transmitter, see
“Switch HART revision mode” on page 4.
Confirm correct Device Driver
Verify that the latest Device Driver (DD/DTM) is loaded on your systems to
ensure proper communication.
Download the latest Device Driver from www.rosemount.com/LevelSoftware.
Rosemount 5300 device revisions and drivers
Tab le 1 provides the information necessary to ensure that you have the correct
Device Driver and documentation for your device.
Table 1. Rosemount 5300 device revisions and files
Switch HART revision mode
If the HART configuration tool is not capable of communicating with HART
Revision 7, the Rosemount 5300 Series will load a Generic Menu with limited
capability. The following procedures will switch the HART revision mode from the
Generic Menu:
1. Manual Setup > Device Information > Identification > Message
To change to HART Revision 5, Enter: “HART5” in the Message field
To change to HART Revision 7, Enter: “HART7” in the Message field
Firmware
Version
1
1. Firmware version is printed on the transmitter head label, e.g. SW 2E0.
Find Device Driver Review Instructions Review Functionality
HART
Universal
Revision
Device
Revision
2
2. Device revision is printed on the transmitter head label, e.g. HART Dev Rev 4.
Manual Document Number Changes to Software
3
3. HART Revision 5 and 7 are selectable.
2F0 and later
74
00809-0100-4530
Rev EA
See footnote 3 for list of
changes.
53
2A2 - 2D2 5 3
00809-0100-4530
Rev DA
N/A