the available pre-defined tank options, or if high volume accuracy is desired.
Choose None if volume calculation is not desired.
Handheld HART Communicator: Fast Key Sequence [2, 1, 5]
F
OUNDATION fieldbus parameters:
TRANSDUCER_1300 > VOL_VOLUME_CALC_METHOD
TRANSDUCER_1300 > VOL_IDEAL_DIAMETER
TRANSDUCER_1300 > VOL_IDEAL_LENGTH
TRANSDUCER_1300 > VOL_VOLUME_OFFSET
Click Next and the following window appears (Analog Output).
10. This step is not applicable for F
OUNDATION fieldbus, the parameters are entered
in the AI-block instead.
Handheld HART Communicator: Fast Key Sequence [2, 1, 6]
For HART communication, choose Primary Variable, PV.
Specify the analog output range by setting the Upper Range Value (20 mA)
and the Lower Range Value (4 mA) to the desired corresponding level values.
The Alarm Mode specifies the output state when a measurement error
occurs. Choose between the following:
High: 21.75 mA (standard) or 22.5 mA (Namur)
Low: 3.75 mA (standard) or 3.6 mA (Namur)
Freeze: the output current is set to the last valid value at the time when the
error occurs.
Click Next.
11. Continue with Steps 2 to 5 in the Guided Setup window:
Step 2: Device specific setup might advise you to make further configurations.
Handheld HART Communicator: Fast Key Sequence [2, 1, 7, 2].
Step 3: Restart the device to make sure all configuration changes are properly
activated.
Step 4: View live values from the device to verify that the transmitter works
correctly.
Step 5: Make a complete backup of the device.
NOTE:
The 4-20 mA range must not include the
upper transition zone, the lower transition
zone, or the upper null zone. See the
Rosemount 5300 Series Reference Manual
(Document No. 00809-0100-4530) for
details.