Quick Installation Guide
IC
1
FOUNDATION fieldbus FISCO model:
Ex ia IIC T4 (-50 °C < T
a
< +60 °C) DIP A20 T
A
69 °C
U
i
= 17.5 V, I
i
= 380 mA, P
i
= 5.32 W, C
i
4.95 nF, L
i
0.1 μH
N3 Type n:
HART model:
Ex nL IIC T4 (-50 °C < T
a
< +70 °C)
U
i
= 30 V, I
i
= 130 mA, P
i
= 1.0 W, C
i
7.26 nF, L
i
0 μH
F
OUNDATION fieldbus model:
Ex nL IIC T4 (-50 °C < T
a
< +60 °C)
U
i
= 30 V, I
i
= 300 mA, P
i
= 1.5 W, C
i
4.95 nF, L
i
0 μH
Japanese certifications
Technology Institution of Industrial Safety (TIIS) approval
Special conditions for safe use (X):
Refer to certificate TC20104 and TC20192.
E4
1
Flameproof:
4-20 mA/HART model:
Transmitter: Ex d [ia] IIC T4x
-20
~
+60 °C
DC 20 - 42.4 V
U
m
= 250 V
U
o
= 22.2 V
I
o
= 177.5 mA
P
o
= 0.985 W
Probe: Ex Ia IIC T4X
F
OUNDATION fieldbus model:
Transmitter: Ex d [ia] IIC T4X
-20
~
+60 °C
DC 16 - 32 V
U
m
= 250 V
U
o
= 22.2 V
I
o
= 177.5 mA
P
o
= 0.985 W
Probe: Ex ia IIC T4X
Installation Drawing: 05300-00548.
Approval valid for HART and F
OUNDATION fieldbus options.
1) Ordering information code for Product Certifications, see the Rosemount 5300 Series Product Data Sheet