Wiringthe9739MVDtransmittertothesensor
Table 4-3 Sensor and transmitter terminal designations
continued
Wire color Sensor terminal Transmitter terminal
Function
Violet
7 7 Temperature +
Gray 8 8
Right pickoff −
White
9 9
Left pickoff −
b. Tighten the screws to hold the wires in place.
c. Ensure integrity of gaskets, grease all O-rings, then replace the junction-box and transmitter housing covers and
tighten all screws, as required.
Sensor and transmitter terminals
Figures 4-5 through 4-7 identify the terminals of the sensors that you can connect to the 9739 MVD transmitter, and
Figure 4-8 identies the 9739 MVD transmitter terminals.
Figure 4-5 ELITE, H-Series, T-Series, and some F-Series sensor terminals
D
I
H
F
E
A
B
C
G
A Violet
B Yellow
C Orange
D Brown
E White
F Green
G Red
H Gray
I Blue
18MicroMotion9739MVDTransmitters