Emerson MMI-20016185 Satellite Radio User Manual


 
Wiringthepowersupplyandgroundingthetransmitter
Important
Follow the plant standards if a separate high-integrity intrinsically safe ground scheme is used.
If national standards or plant standards are not in effect, adhere to the following guidelines for grounding:
Use copper wire, 14 AWG (2.5 mm
2
) or larger wire size.
Keep all ground leads as short as possible, less than 1 Ω impedance.
You can use the internal power ground terminal or the external ground terminal to ground the transmitter. Directly
connect the ground lead from the terminal to Earth ground.
To achieve potential equalization and comply with ATEX standards for hazardous area installations, connect the
external ground terminal to the appropriate ground terminals within the hazardous area, using a potential-equalizing
line.
Procedure
Refer to Figure 3-2 for information on grounding the transmitter.
Figure 3-2 Grounding the 9739 MVD transmitter
A Power ground terminal
B External ground terminal
C Earth ground
10MicroMotion9739MVDTransmitters