Fluke Rev. 4 Speaker User Manual


 
Documenting Process Calibrator
Specifications
89
DC Current Measurement
RANGE RESOLUTION
MODEL 702
% OF READING +% OF FULL SCALE
MODEL 701
% OF READING +% OF FULL SCALE
1 YEAR 2 YEAR 1 YEAR 2 YEAR
30 mA 1 µA 0.025% + 0.025% 0.025% + 0.03% 0.025% +0.025% 0.025% + 0.03%
110 mA 10 µA 0.05% + 0.05% 0.07% + 0.07% 0.05% + 0.05% 0.1% + 0.07%
Temperature Coefficient: (0.001% of rdg. + 0.002% f.s.)/°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.01% f.s./(Common Mode Volt)
Maximum Input Voltage: 30V dc
Resistance Measurement
RANGE RESOLUTION
MODEL 702
% OF READING + Ohms
MODEL 701
% OF READING + Ohms
1 YEAR 2 YEAR 1 YEAR 2 YEAR
11 0.001 0.05% + 0.05 0.075% + 0.05 0.1% + 0.05 0.15% + 0.05
110 0.01 0.05% + 0.05 0.075% + 0.05 0.1% + 0.05 0.15% + 0.05
1.1 k 0.1 0.05% + 0.5 0.075% + 0.5 0.1% + 0.5 0.15% + 0.5
11 k 1 0.1% + 10 0.1% + 10 0.1% + 10 0.15% + 10
Temperature Coefficient: (0.01% f.s. + 2 mΩ) /°C in the ranges -10 to 18°C and 28 to 50°C
Common Mode Error: 0.005% f.s/(Common Mode Volt)
Maximum Input Voltage: 30V dc