Fluke Rev. 4 Speaker User Manual


 
Documenting Process Calibrator
Setting Up the Calibrator
23
Automatically Saving Battery Life
An optional Auto Battery Save feature turns the
calibrator off after a selected idle time. The default
setting for Auto Battery Save is Off. When Auto
Battery Save is On, the Esymbol shows in the
upper right corner of the display. The setting is
preserved after you turn off the power. Auto Battery
Save works the same on battery power and when
using the battery eliminator. Turn on the Auto
Battery Save feature as follows:
1. Press s.
2. Press d to highlight Off following Auto Battery
Save.
3. Press e or the Choices softkey.
4. Press u to highlight On, then press e.
5. To accept the timeout period shown on the
display, you can finish here. Press Done to exit
s mode and do not go on to step 6.
6. To change the timeout period, press d to
highlight the timeout period following Battery
Save Timeout.
7. Press e or the Choices softkey.
8. Enter your choice of timeout period in minutes
(accepted range: 1 to 120 minutes).
9. Press the Done softkey.
10. Press the Done softkey or s to exit s
mode.
Using the Optional Battery Eliminator
CAUTION
To avoid damage to the calibrator, use
only Fluke Model BE9005 Series Battery
Eliminator, available from your
distributor.
Where ac power is available, you can use the
optional Fluke Model BE9005 Battery Eliminator to
conserve battery power. When the battery eliminator
is used, the battery is internally disconnected, and
can be removed from the calibrator. The calibrator
operates the same as it does under battery power,
however, the specifications for the 701/702 only
apply under battery operation. The battery eliminator
does not charge the battery.