Fluke Rev. 4 Speaker User Manual


 
701/702
Users Manual
88
AC Voltage Measurement
FREQUENCY RANGE
MODEL 701 AND 702
(% OF READING + NUMBER OF COUNTS)
1 YEAR 2 YEAR
20 to 40 Hz 2% + 10 2% + 10
40 to 500 Hz 0.5% + 5 0.5% + 5
500 Hz to 1 kHz 2% + 10 2% + 10
1 kHz to 5 kHz 10% + 20 10% + 20
Specifications apply for 10% to 100% of voltage range
Ranges : 1.1000 V, 11.000 V, 110.00 V, 300.0 V rms
Resolution: 11.000 counts in all ranges except 300V; 3,000 counts on 300V range.
Input Impedance: 5 M and <100 pF
Temperature Coefficient: 10% of spec/°C in the ranges -10 to 18°C and 28 to 50°C
Input Coupling: ac
Maximum Input Voltage: 300 V rms
For frequencies >1 kHz, the minimum voltage is 0.5 V