Fluke Rev. 4 Speaker User Manual


 
Documenting Process Calibrator
Using Measure Mode
37
To measure temperature using an RTD input,
proceed as follows:
1. If necessary, press M for MEASURE mode.
2. Press ttwice so that Select RTD Type
shows.
3. Press u or d to select the desired RTD type.
4. Press e.
5. Press u or d to select a 2-, 3-, or 4- wire
connection.
6. Attach the RTD to input jacks as the display or
Figure 13 shows. Use the supplied jumper
between the mA
RTD MEAS low jack and the
V MEAS low jack as shown if you are using a 3-
wire connection.
7. Press e.
CAUTION
Do not force a dual banana plug between
any two jacks in the horizontal
orientation. (See the Figure.) Doing so
will damage the jacks. Use the supplied
jumper wire when needed for RTD
measurements. You can use a dual
banana plug in the vertical orientation.