Fluke Rev. 4 Speaker User Manual


 
701/702
Users Manual
40
Measurements in Percent of Scale
To display measurements as a percent of scale as
shown below, proceed as follows:
1. If necessary, press M for MEASURE mode.
2. Select a measurement function (m, v, q,
t, or p) as previously described.
3. Press the Scale softkey.
4. Use the numeric keypad to enter the 0% of scale
value.
5. Press e.
6. Use the numeric keypad to enter the 100% of
scale value.
7. Press e.
8. Press the Done softkey.
Percent of scale remains in effect until you press the
Scale softkey again or change to another
measurement function.
Damping Your Measurements
The calibrator normally applies a software filter to
dampen measurements in all functions except
continuity. The specifications assume that damping is
turned on. The damping method is a running average
of the last several measurements. Fluke
recommends that you leave damping on. Turning
damping off may be useful when measurement
response is more important than accuracy or noise
reduction. If you want to turn off damping, press the
More Choices softkey twice, then press the
Dampen softkey so that Off appears. Press
Dampen again to turn damping back on. The default
state is On.