value is shown in the top left of the measure
window.
15. The calibrator moves on to the remaining set of
points. For temperature and electrical parameter
calibration, the points are done automatically. If
you are sourcing pressure, the calibrator pauses
at each step for you to adjust the pressure
source. When the tests are complete, an error
summary table, such as the following is
displayed.
16. In the results summary test, failures are
highlighted. An adjustment is required in this
example because three tests show failures. The
failures were outside the ±0.5% tolerance that
we selected.
17.
Either press the Done softkey to save the data,
or the Abort softkey to delete the data and start
over. You have to go back to As Found to
restart the calibration. You can see the saved
data entry and recall the table for later viewing
through the Review Memory softkey during
normal operation. With Model 702 you can
upload this data to a PC using PMLink software.
See the
PMLink Manual
for instructions.
Adjusting the Transmitter
Proceed as follows to make the calibration
adjustments to the transmitter. (Always refer to the
transmitter manufacturer’s instructions to locate the
adjustment controls and connection points for your
transmitter).
1.
Press the Done softkey while viewing the results
summary.
2.
Press the Adjust softkey. The calibrator sources
0% of span (100°C in this example) and displays
the following softkeys: