Table 1-2. 5700A Series II DC Voltage Specifications: 99 % and 95 % Confidence Levels
5700A
99%
99 % Confidence Level
Range Resolution
Absolute Uncertainty
± 5 °C from calibration temperature
For fields strengths >1 V/m but 3 V/m,
add 0.01 % of range.
Relative Uncertainty
± 1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
± (ppm output + µV) ± (ppm output + µV)
220 mV 10 nV 6.5 + .75 7 + .75 08 + .75 09 + .8 2.5 + .5 04 + .5
2.2 V 100 nV 3.5 + 1.2 6 + 1.2 07 + 1.2 08 + 1.2 2.5 + 1.2 04 + 1.2
11 V 1 µV 3.5 + 3 5 + 4 07 + 4 08 + 4 1.5 + 3 3.5 + 4
22 V 1 µV 3.5 + 6 5 + 8 07 + 8 08 + 8 1.5 + 6 3.5 + 8
220 V 10 µV 05 + 100 6 + 100 08 + 100 09 + 100 2.5 + 100 04 + 100
1100 V 100 µV 07 + 600 8 + 600 10 + 600 11 + 600 03 + 600 4.5 + 600
5700A
95%
95 % Confidence Level
Range Resolution
Absolute Uncertainty
± 5 °C from calibration temperature
For fields strengths >1 V/m but 3 V/m,
add 0.01 % of range.
Relative Uncertainty
± 1 °C
24 Hours 90 Days 180 Days 1 Year 24 Hours 90 Days
± (ppm output + µV) ± (ppm output + µV)
220 mV 10 nV 5.5 + 0.6 6 + 0.6 7 + 0.6 8 + 0.6 02 + 0.4 3.5 + 0.4
2.2 V 100 nV 3.5 + 1 5 + 1 6 + 1 7 + 1 02 + 1 3.5 + 1
11 V 1 mV 03 + 3.5 4 + 3.5 6 + 3.5 7 + 3.5 1.2 + 3 03 + 3.5
22 V 1 mV 03 + 6.5 4 + 6.5 6 + 6.5 7 + 6.5 1.2 + 6 03 + 7
220 V 10 mV 04 + 80 5 + 80 7 + 80 8 + 80 02 + 80 3.5 + 80
1100 V 100 mV 06 + 500 7 + 500 8 + 500 9 + 500 2.4 + 500 04 + 500