5725A
Instruction Manual
6-4
100V
110V
115V
120V
S2 S3 S4
VOLTAGE
SELECTION
FUSE
F8A 250V
(FAST)
200V
220V
230V
240V
S2 S3 S4
F4A 250V
(FAST)
JOHN FLUKE MFG. CO., INC.
MADE IN U.S.A.
PATENTS PENDING
NO INTERNAL USER SERVICEABLE
PARTS REFER SERVICE TO
QUALIFIED SERVICE PERSONNEL
TO CLEAN FILTER
REMOVE FROM INSTRUMENT
AND FLUSH WITH WARM
SOAPY WATER
5700A CALIBRATOR
20V PK
MAX
CHASSIS
GROUND
FUSE
WARNING
GROUNDING
CONNECTOR IN POWER CORD
MUST BE CONNECTED TO
ENSURE PROTECTION FROM
ELECTRONIC SHOCK.
47-63 Hz
750VA MAX
CAUTION FOR FIRE PROTECTION
REPLACE ONLY WITH A 250V FUSE
OF INDICATED RATING.
S4S3S2
11A MAX
HI
LO
CALIBRATION
ENABLE NORMAL
CURRENT
OUTPUT
!
!
!
aq13f.eps
Figure 6-1. Accessing the Fuse
6-3. Cleaning the Air Filter
Caution
Damage caused by overheating may occur if the area around
the fan is restricted, the intake air is too warm, or the air filter
becomes clogged.
The air filter must be removed and cleaned every 30 days, or more frequently if the
calibrator is operated in a dusty environment. The air filter is accessible from the rear
panel of the calibrator.
To clean the air filter, refer to Figure 6-2 and proceed as follows:
1. Turn off the POWER switch and disconnect the line power cord from ac line power.
2. Squeeze together the sides of the air filter housing and pull it off.
3. Wash the filter element and housing in soapy water.
4. Rinse the filter element and housing in fresh running water.
5. Shake out excess water, then allow the filter element to dry thoroughly before
reinstalling it.
6. Snap the filter housing back into place.