3-14
ADVISE_MASK
The ADVISE_MASK parameter will mask any of the failed conditions listed
in ADVISE_ENABLED. A bit on means the condition is masked out from
alarming and will not be reported.
ADVISE_PRI
ADVISE_PRI designates the alarming priority of the ADVISE_ALM, see
Table 3-4 on page 3-11. The default is 0 and the recommended values are
1 or 2.
ADVISE_ACTIVE
The ADVISE_ACTIVE parameter displays which of the advisories is
active. Only the advisory with the highest priority will be displayed. This
priority is not the same as the ADVISE_PRI parameter described above.
This priority is hard coded within the device and is not user configurable.
ADVISE_ALM
ADVISE_ALM is an alarm indicating advisory alarms. These conditions do
not have a direct impact on the process or device integrity.
Recommended Actions
for PlantWeb Alerts
RECOMMENDED_ACTION
The RECOMMENDED_ACTION parameter displays a text string that will give
a recommended course of action to take based on which type and which
specific event of the PlantWeb alerts are active.
Table 3-7.
RB.RECOMMENDED_ACTION
Alarm Type Active Event Recommended Action
None None No action is required.
Advisory PWA Simulate Active Disable simulation to return to process
monitoring.
Advisory Excessive Deviation
Advisory Excessive Rate of
Change
Maintenance CJC Degraded If T/C sensors are being used, restart the
device. If condition persists, replace the
device.
Maintenance Body Temperature Out
of Range
Verify the ambient temperature is within
operating limits.
Maintenance Sensor 1 Degraded Confirm the operating range of Sensor 1
and/or verify the sensor connection and
device environment.
Maintenance Sensor 2 Degraded Confirm the operating range of Sensor 2
and/or verify the sensor connection and
device environment.
Maintenance Sensor 3 Degraded Confirm the operating range of Sensor 3
and/or verify the sensor connection and
device environment.
Maintenance Sensor 4 Degraded Confirm the operating range of Sensor 4
and/or verify the sensor connection and
device environment.
Maintenance Sensor 5 Degraded Confirm the operating range of Sensor 5
and/or verify the sensor connection and
device environment.
Maintenance Sensor 6 Degraded Confirm the operating range of Sensor 6
and/or verify the sensor connection and
device environment.
Maintenance Sensor 7 Degraded Conform the operating range of Sensor 7
and/or verify the sensor connection and
device environment.