Anritsu ME7840A Stereo Amplifier User Manual


 
5-6 S-PARAMETER
TEST: K FACTOR
This test describes the K Factors measurement. This measurement is
described in Chapter 1, paragraph 1-10.
Procedure.
Prepare the ME7840A as shown in Chapter 3.
Enter the appropriate test values in the displayed
User Input screen. The input fields are described
below: The inputs required for this test are de
-
scribed below.
n
S-Parameter: Drop down options. K Factor for
this measurement.
n
Data Points: Drop down options - number of data
points in frequency sweep.
n
Frequency 1 (MHz): Lower frequency or start fre-
quency.
n
Frequency 2 (MHz): Upper frequency or stop fre-
quency.
n
Input Power (dBm):Input power level, in dBm.
n
IF Bandwidth (Hz):Intermediate frequency band-
width for test, in Hertz.
n
Test Set Setting: Byte value for setting test set
attenuation and source switch (below).
OPERATIONS, MEASUREMENT S-PARAMETER TEST: K FACTOR
ME7840A OMM 5-7
Internal Source Value External Source Value
0dB 8 0dB 0
10 dB 9 10 dB 1
20 dB 10 20 dB 2
30 dB 11 30 dB 3
40 dB 12 40 dB 4
50 dB 13 50 dB 5
60 dB 14 60 dB 6
70 dB 15 70 dB 7