Anritsu ME7840A Stereo Amplifier User Manual


 
A-7 HOT S
22
.TXT The HotS
22
.txt file is shown in Figure A-3.
CAL SPECIFICATION FILES HOT S
22
.TXT
ME7840A OMM A-13
ndut,1, //Number of DUTs
type,sp, //S Parameter calibration
SP_START
ncal,1, //Number of S-Param calibrations
npts,201, //Number of points (default for all)
con,FN,FN, //Connector Type Options MN,FN,MK,FK,MS,FS
if,3000, //IF default
avgs,0, //averages default
pow,-5, //power level default
switch,8, //switch setting
CAL,1, caltype,RR, //12 = 12 Term, RR = Relection only Reverse, RF =
Relection only Forward
name,Hots22, //File name root (append _1 for DUT1, _2 for DUT2, etc
hot,923, //hot 0 means turn on 2nd source during cal (Hot S22
only),
//use value as offset (KHz)
pow2,-5, //2nd source power in Hot S22
f1,800, //startf DUT1 DUT2 DUT3
f2,1000, //stopf DUT1 DUT2 DUT3
SP_END
END
Figure A-3. Hot S
22
Calibration Specification Program File