4 To configure your analyzer for a keyboard other than U.S. English, press
[
System Utility
][
keyboard type
]. Then press the appropriate softkey to select
the language.
Configuring your analyzer to use a keyboard other than U.S. English only
ensures that the analyzer recognizes the proper keys for that particular
keyboard. Configuring your analyzer to use another keyboard does not localize
the on-screen annotation or the analyzer’s online HELP facility.
To connect the optional minimum loss pad
The minimum loss pad (option 1D7) provides a 50 ohm matched impedance to
the analyzer and a 75 Ω matched impedance to the device under test.
1 Connect the minumum loss pad to the RF section’s INPUT or SOURCE
connector.
2
Connect a 75 Ω cable between the minimum loss pad and the device under test.
Use either a 75 Ω type-N cable or the supplied 75 Ω type-N(m)-to-BNC(f)
adapter and a 75 Ω BNC cable.
Caution
Do NOT connect a 50 Ω cable or adapter to the 75 Ω minimum loss pad. The
center pin is larger in a 50 Ω type-N connector than in a 75 Ω type-N connector.
Connecting a 50 Ω type-N connector to the 75 Ω minimum loss pad will damage
the 75 Ω minimum loss pad.
Preparing the Analyzer for Use
7-18