To do the incoming inspection
The analyzer was carefully inspected both mechanically and electrically before
shipment. It should be free of marks or scratches, and it should meet its
published specifications upon receipt.
1 Inspect the analyzer for physical damage incurred in transit. If the analyzer
was damaged in transit, do the following:
Save all packing materials.
File a claim with the carrier.
Call your Agilent Technologies sales and service office.
Warning If the analyzer is mechanically damaged, the integrity of the protective
earth ground may be interrupted. Do not connect the analyzer to
power if it is damaged.
2 Check that the line-voltage selector switches are set for the local line voltage.
Theline-voltageselectorswitchesaresetatthefactorytomatchthemost
commonly used line voltage in the country of destination. To check or change
the line-voltage selector switches, see ‘’To change the IF section’s line-voltage
switch’’ and ‘’To change the RF section’s line-voltage switch.’’
3 Check that the correct line fuses are installed in the fuse holders.
Thefusesareinstalledatthefactoryforthemostcommonlyusedlinevoltagein
the country of destination. There is one line fuse in the IF section and one line
fuse in the RF section. To determine if the correct line fuses are installed, see
‘’To change the IF section’s fuse’’ and ‘’To change the RF section’s fuse.’’
4 Connect the IF section to the RF section.
For instructions on connecting the sections, see ‘’To connect the sections.’’
5 Using the supplied power cords, plug the analyzer’s IF section and RF section
into appropriate receptacles.
The analyzer is shipped with two three-conductor power cords that ground the
analyzer when plugged into appropriate receptacles. The type of power cable
plug shipped with each analyzer depends on the country of destination.
Preparing the Analyzer for Use