Agilent Technologies 89441A Stereo System User Manual


 
If the IF section will not power up
q Check that the power cord is connected to the IF section and to a live power
source.
q Check that the front-panel switch is on ( l ).
q Check that the voltage selector switch is set properly.
See ‘’To change the IF section’s line-voltage switch’’ on page 7-10.
q Check that the fuse is good.
See ‘’To change the IF section’s fuse’’ on page 7-12.
q Check that the IF section’s air circulation is not blocked.
Cooling air enters the IF section through the rear panel and exhausts through
both sides. If the IF section’s air circulation is blocked, the IF section powers
down to prevent damage from excessive temperatures. The IF section remains
off until it cools down and its power switch is set to off (
O )thentoon(l ).
q Obtain service, if necessary. See ‘’Need Assistance?’’ at the end of this guide.
Preparing the Analyzer for Use