6 Set the RF section’s rear panel and front panel power switches to on.
Press the ‘’
l‘’ symbol end of the rocker-switches located on the lower right of the
rear panel and on the lower left of the front panel. The RF section provides
standby power for the high precision frequency reference. The rear-panel line
switch interrupts all power including standby power when you press the ‘’
O‘’
symbol end of the switch. The front-panel power switch interrupts all power
except standby power when you press the ‘’O
I
‘’ symbol end of the switch.
7 Set the IF section’s power switch to on.
Press the ‘’
l‘’ symbol end of the rocker-switch located on the lower left of the
front panel. The analyzer requires about 30 seconds to complete its power-on
routine.
8 Test the electrical performance of the analyzer using the operation verification
or the performance tests in chapter 2, ‘’Verifying Specifications’’ in the
Installation and Verification Guide.
The operation verification tests verify the basic operating integrity of the
analyzer; these tests take about 2.5 hours to complete and are a subset of the
performance tests. The performance tests verify that the analyzer meets all the
performance specifications; these tests take about 5 hours to complete.