44 Model IFT9701 Transmitter Instruction Manual
Configuration with a HART Communicator continued
6. Multiply the meter factor from Step 5 by the first five digits of the current flow calibration
factor. This is the first five digits of the new flow calibration factor.
7. To enter the new flow calibration factor into flowmeter memory, refer to the menu below and
follow these steps:
a. At the online menu, choose 4 (detailed setup).
b. At the detailed setup menu, choose 1 (characterize sensor).
c. At the characterize sensor menu, choose 1 (flow calibration factor).
d. Type in the new flow calibration factor, then press F4 (enter):
e. For the first five digits and first decimal point, use the value determined in Step 6.
f. For the last three digits and second decimal point, use the last three digits and second
decimal point from the flow calibration factor listed on the flowmeter serial number tag.
g. When the display returns to the characterize sensor menu, press F2 (send) to send the
factor to the flowmeter memory.
8. To verify the accuracy of the new flow calibration factor, repeat Step 4. The amount of fluid
indicated by the mass totalizer should equal the weighed amount of fluid in the batch, within
accuracy specifications provided by Micro Motion for the flowmeter.
6.4 Characterization
Characterization involves entering a density calibration factor that accurately represents the
sensitivity of the flowmeter to fluid density.
To characterize the flowmeter for density measurement, refer to the menu below and follow these
steps:
1. At the online menu, choose 4 (detailed setup).
2. At the detailed setup menu, choose 1 (characterize sensor).
3. At the characterize sensor menu, choose 2 (density calibration factor).
4. Enter the 13-digit density calibration factor listed on the flowmeter serial number tag. The
density calibration factor can also be found on the certificate that is shipped with the meter.
5. When the display returns to the characterize sensor menu, press F2 (send) to send the factor to
the flowmeter memory.
First 5 digits of new FloCal factor
1 Process variables
2 Diag/service
3 Basic setup
4 DETAILED SETUP
5Review
1 CHARIZE SENSOR
2 Config fld dev var
3 Config outputs
4 Device information
1 FLOW CAL
2 Density cal factr
1 Process variables
2 Diag/service
3 Basic setup
4 DETAILED SETUP
5Review
1 Flow cal
2 DENSITY CAL FACTR1 CHARIZE SENSOR
2 Config fld dev var
3 Config outputs
4 Device information