Emerson IFT9701 Satellite Radio User Manual


 
36 Model IFT9701 Transmitter Instruction Manual
Configuration with a HART Communicator continued
6.2 Configuration parameters
Use the basic setup menu to perform the following tasks:
Assigning a HART tag to the flowmeter
Changing measurement units for the mass or volume flow rate
Setting range values for the milliamp output
Scaling the pulse output
Use the detailed setup menu to perform the following tasks:
Changing measurement units for density or temperature
Changing the flow cutoff value
Changing the internal damping value
Changing the flow direction parameter
6.2.1 HART tag
The HART tag consists of up to eight characters that identify the flowmeter when it communicates
with other devices in a HART multidrop network.
To assign a HART tag to a flowmeter, use the menu below and follow these steps:
1. At the online menu, choose 3 (basic setup).
2. At the basic setup menu, choose 1 (tag).
3. Enter the desired tag of up to eight characters, then press F4 (enter). The tag can include
spaces and periods.
6.2.2 Measurement units
The meter can use one mass flow unit or one volume flow unit. The meter can also provide density
and temperature indication to a HART Communicator.
To select a measurement unit for mass flow or volume flow, use the menu below and follow these
steps:
1. At the online menu, choose 3 (basic setup).
2. At the basic setup menu, choose 2 (primary variable unit).
3. Use the down arrow (Ø) or up arrow (×) to select the desired measurement unit, then press F4
(enter). Available measurement units for mass flow and volume flow are listed in Table 6-1.
1 Process variables
2 Diag/service
3 BASIC SETUP
4 Detailed setup
5Review
1TAG
2PV unit
3 Analog range vals
4 Freq factr
1 Process variables
2 Diag/Service
3 BASIC SETUP
4 Detailed setup
5Review
1Tag
2 PV UNIT
3 Analog range vals
4 Freq factr