Kodak 1H9659 Radio User Manual


 
Calibration Utility for MICROSOFT WINDOWS
1–32
September 1997 Part No. 1H9659
6. Click on the
Print
button.
You return to the Calibration Document window. The calibration test
print is printed on the specified printer.
7. View the calibration test print and determine which cluster appears
the most neutral.
If the Visual Gray Balance most-neutral cluster number on the
test print ends in 25, the printer is already calibrated. You should
now perform a Visual Linearity calibration.
If the Visual Gray Balance number on the test print does not
end in 25, you must recalibrate the printer to the most neutral
cluster number.
NOTE: If you enter a value that ends in 25, the calibration table does
not change.
8. Click on the
Input
button.