Agilent Technologies 6632B Stereo System User Manual


 
2 - Verification and Performance Tests
22
Model Agilent 6634B Report No _______________ Date __________________
Test Description Minimum
Specs.
Results* Maximum
Specs.
Measurement
Uncertainty
Constant Voltage Tests
Voltage Programming and Readback
Low Voltage (0V) Vout
Front Panel Display Readback
High Voltage (Full Scale) Vout
Front Panel Display Readback
50 mV
Vout 12 mV
99.9 V
Vout 42 mV
__________
__________
__________
__________
+ 50 mV
Vout + 12 mV
100.1 V
Vout + 42 mV
2.1 µV
2.1 µV
1.4 mV
1.4 mV
Load Effect
5 mV
__________ + 5 mV
60 µV
Source Effect
1 mV
__________ + 1 mV
60 µV
PARD (Ripple and Noise)
Peak-to-Peak
RMS
0 mV
0 mV
__________
__________
+ 3 mV
+ 0.5 mV
872 µV
50 µV
Transient Response
Time in 100 µs
0 mV __________ + 100 mV 15 mV
Constant Current Tests
Current Programming and Readback
Low current (0A) Iout
Readback Accuracy @ Iout
High Current (Full Scale) Iout
Readback Accuracy @ Iout
0.5 mA
Iout 0.25 mA
0.999 A
Iout 2.3 mA
__________
__________
__________
__________
+ 0.5 mA
Iout + 0.25 mA
1.001 A
Iout + 2.3 mA
15.1 µA
15.1 µA
128.8 µA
128.8 µA
Current Sink Readback
Isink 2.9 mA
__________
Isink + 2.9 mA
128.8 µA
20 mA Range Current Readback
Readback Accuracy @ 0 A
Readback Accuracy @ + 20 mA
Readback Accuracy @ 20 mA
2.5 µA
Iout 22.5 µA
Iout 22.5 µA
__________
__________
__________
+ 2.5 µA
Iout + 22.5 µA
Iout + 22.5 µA
0.1 µA
1.7 µA
1.7 µA
PARD (Current Ripple and Noise)
RMS
0 mA __________ + 2.0 mA
250 µA
Load Effect
1.0 mA
__________ + 1.0 mA
1 µA
Source Effect
0.25 mA
__________ + 0.25 mA
1 µA
* Enter your test results in this column