Agilent Technologies 6632B Stereo System User Manual


 
1 - Introduction
10
Revisions
Manual Revisions
This manual was written for dc power supplies that have the same manufacturing dates (the first four digits) as those
listed on the title page and whose unique identification number (the last four digits) are equal to or higher than those
listed in the title page.
NOTE: If the first four digits of the serial number of your unit are higher than those shown in the title
page, your unit was made after the publication of this manual and may have hardware or firmware
differences not covered in this manual. If they are significant to the operation and/or servicing of
the dc power supply, those differences are documented in one or more Manual Change sheets
included with this manual.
Firmware Revisions
You can obtain the firmware revision number by either reading the integrated circuit label, or query the dc power
supply using the GPIB *IDN?' query command (See Chapter 3, ROM Upgrade).
Electrostatic Discharge
CAUTION: The dc power supply has components that can be damaged by ESD (electrostatic discharge).
Failure to observe standard antistatic practices can result in serious degradation of performance,
even when an actual failure does not occur.
When working on the dc power supply, observe all standard, antistatic work practices. These include, but are not
limited to:
a Working at a static-free station such as a table covered with static-dissipative laminate or with a conductive
table mat (Agilent P/N 9300-0797, or equivalent).
a Using a conductive wrist strap, such as Agilent P/N 9300-0969 or 9300-0970.
a Grounding all metal equipment at the station to a single common ground.
a Connecting low-impedance test equipment to static-sensitive components only when those
components have power applied to them.
a Removing power from the dc power supply before removing or installing printed circuit boards.