Agilent Technologies 83430-90011 Satellite Radio User Manual


 
5-37
Servicing
Adjustment Procedures
Table 5-6. Required Test Equipment
Equipment
Agilent
Recommended
Model
Critical Specifications
Digital voltage meter 34401A
150 Vdc range. 1 mV accuracy.
Digitizing oscilloscope 83480A with
83483A and
83485A
>10 GHz bandwidth
10 ps/div amplitude scale
500 MHz triggering, optical input with built-in SDH filter.
Voltage and time histograms. Built-in SONET/SDH eye mask
conformance test.
Pattern generator 70841B
0.1 to 3.3 Gb/s, 2
23
1 PRBS.
250 mV p-p to 1200 mV p-p output.
Rise/fall times
100 ps.
Clock/32 trigger.
50
output.
Clock source 70311A
16.1 MHz to 3.3 GHz, 50
, output.
<120 dBc noise
Sinewave or Squarewave, 500 mV p-p to 1500 mV p-p
Mainframe 70001A Specific mainframe for pattern generator and clock source.
Optical power meter 1310/1550 nm calibration. FC connector
0.1 dB resolution, 0.2 dB accuracy
60 to +20 dBm range