Agilent Technologies 83430-90011 Satellite Radio User Manual


 
5-16
Servicing
Performance Tests
2
Set the pattern generator to the following settings. For step-by-step
instructions on setting these values, refer to Configuring the pattern
generator on page 5-35.
data pattern. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
23
1
data amplitude (ECL). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 850 mV p-p
data output high level (ECL). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 900 mV
data termination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 V
trigger mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .CLOCK / 23
trigger pattern. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23 zeros
clock rate. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.48832 GHz
clock amplitude. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 850 mV
3
Press the Agilent 83430As front-panel
SELECT
key repeatedly until the
DIGITAL
IN AC COUPLED
light is on.
4
While turning the Agilent 83430As front-panel
DIGITAL THRESHOLD
knob fully
clockwise and fully counterclockwise, observe the following items:
The eye is not squelched for any setting of the knob.
The eye diagrams zero crossing varies as the knob is turned.
The eye diagrams zero crossing should be approximately in the middle
when the knob is set to its center position.
5
Disconnect the modulation signal from the
DIGITAL IN AC COUPLED
connector, and
connect it to the
DIGITAL IN DC COUPLED
connector.
6
Press the Agilent 83430As front-panel
SELECT
key so that the
DIGITAL IN DC
COUPLED
light is on.
7
While turning the Agilent 83430As front-panel
DIGITAL THRESHOLD
knob fully
clockwise and fully counterclockwise, observe the following items:
The eye is not squelched for any setting of the knob.
The eye diagrams zero crossing varies as the knob is turned.