Agilent Technologies 83430-90011 Satellite Radio User Manual


 
5-13
Servicing
Performance Tests
Performance Tests
The procedures in this section test the Agilent 83430As performance using
the specifications listed in Chapter 3, Specifications and Regulatory Informa-
tion as the performance standard. All of the tests are done manually without
the aid of a computer. None of these tests require access to the interior of the
instrument. Allow the Agilent 83430A to warm up for 15 minutes before doing
any of the performance tests.
If the instrument fails any performance test, perform the adjustment proce-
dures located in Adjustment Procedures on page 5-36.
The following performance tests are included in this section:
Test 1. Digital Threshold 5-15
Test 2. Frequency Response (ANALOG IN) 5-17
Test 3. Electrical Return Loss (ANALOG IN) 5-20
Test 4. Electrical Return Loss (DIGITAL IN) 5-22
Test 5. Eye Quality (DIGITAL IN) 5-24
Test 6. STM-16/OC-48 Conformance 5-26
Test 7. OC-1 Conformance 5-28
Test 8. Output Power 5-31
Test 9. Output Wavelength 5-33
Calibration Cycle
This instrument requires periodic verification of performance. The instrument
should have a complete verification of specifications at least once every two
years.
CAUTION
Option 022 instruments have an angled-fiber output. Be sure to use an angled-
fiber patchcord during testing. If you do not have an angled-fiber patchcord
available, you can purchase the required accessories from Agilent
Technologies. These include the Agilent 81000SI DIN connector interface and
the Agilent 81113PC DIN 4108 (angled) to Super-PC patchcord.