Emerson 2700 Satellite Radio User Manual


 
22 Micro Motion
®
Model 1700 and 2700 Transmitters
Wiring the Transmitter to the Sensor
3.4 Wiring for 9-wire remote installations
If you chose a 9-wire remote installation (see Figure 2-1), a 9-wire cable must be used to connect the
junction box on the sensor to the core processor on the transmitter/core processor assembly.
To connect the cable, follow the steps below:
1. Refer to Micro Motion’s 9-Wire Flowmeter Cable Preparation and Installation Guide for
instructions on cable shielding and preparation:
At the sensor end, follow the instructions for your cable type.
At the transmitter end, follow the instructions for your cable type with an MVD
transmitter.
2. To connect the wires, refer to Micro Motion’s 9-Wire Flowmeter Cable Preparation and
Installation Guide and follow the instructions for your sensor with an MVD transmitter.
Additional information for connecting the wires at the transmitter is provided below:
a. Identify the components shown in Figure 2-6.
b. Remove the end-cap.
c. Insert the 9-wire cable through the conduit opening.
d. Connect the wires to the plugs supplied with the transmitter.
e. Insert the plugs into the sockets inside the lower conduit ring. See Figure 3-4.
Figure 3-4 9-wire cable between sensor and core processor (on transmitter)
3. Ground the cable.
CAUTION
Allowing the shield drain wires to contact the sensor junction box can cause
flowmeter errors.
Do not allow the shield drain wires to contact the sensor junction box.
Brown
Red
Green
White
Blue
Gray
Orange
Violet
Ye l l o w
Black
(Drains from all
wire sets)
Plug and
socket
Mounting screw
Blue
Gray
Orange
Red
Green
White
Brown
Violet
Yellow
Ground screw
Black
9-wire cable from sensor Core processor (on transmitter)