Signal Output
Signal attenuation
75
From Table 5•3,note that for a TCD,
RANGE 2 ()
= 0 is suitable for
virtually all applications since the entire linear output range of the
detector is included. Likewise,
RANGE 2 ()
settings from 0 through 5 cover
the entire useful output range for an ECD. Only an FID or NPD may
require use of the higher
RANGE 2 ()
settings.
Displaying current
RANGE 2 ()
/
ATTN 2 ()
setpoints
Current setpoint value for
RANGE 2 ()
or
ATTN 2 ()
is displayed by
pressing the appropriate signal channel key (
SIG 1
or
SIG 2
), followed
by either
RANGE 2 ()
or
ATTN 2 ()
(or simply press
RANGE 2 ()
or
ATTN 2 ()
alone, if the desired signal channel is already displayed).
For example, the following key sequence displays current setpoint for
RANGE 2 ()
for the
SIG 2
output channel:
SIG 2
RANGE 2 ()
Typical displays that occur are shown in Figure 5•4.
ACTUAL SETPOINT
SIG 2 NOT INSTALLED
ACTUAL SETPOINT
SIGNAL 1 RANGE 10
SIGNAL 1 ATTN 10
ACTUAL SETPOINT
SIGNAL 1 ATTN 10 OFF
ACTUAL SETPOINT
Typical Signal Channel Displays
Figure 5-4