A SERVICE OF

logo

Test Sample Chromatograms
Test sample chromatograms
235
Detector Type TCDw/MUG
Temp 300 DEGREES C
Inlet Type Ded
On-Col
Oven Track On
Temp N/A C
Operating Mode N/A
Purge Time On N/A min
Purge Time Off N/A
min
Oven Temp Programmed (1 ramp)
Init Temp 60
DEGREES C
Init Time 0.5
min
Ramp
Rate 20
Fin Temp 180
Fin Time 1
Flow Param (EPP)
Constant Flow Off
Range 0
COLUMN:
Part No. 19095S(#100)
Dimensions 530 MID; 5 M
Sta Phase Methyl
Silicone
4.23
4.82
5.38
FLOW RATES
Carrier (He) 20 +/• 1 ml/min
Hydrogen N/A ml/min
Air N/A ml/min
Makeup (He) 5+/•1
ml/min
Split Vent N/A
ml/min
Septum Purge 5+/•1
ml/min
Reference (He) 37+/•1
ml/min
SAMPLE:
Type FID Sample
Inj Volume 3
l
Part No. 18710•60170
Composition 0.03%(V/V)each
C14,C15,C16 normal paraffin
hydrocarbons in hexane.
C14
C15
C16
HP 5890 Test Sample Operating
Conditions
TCD-On-Column Capillary Inlet
Figure 10-11.