Test Sample Chromatograms
Test sample chromatograms
225
Test sample chromatograms
Detector Type FID (or FIDw/MUG)
Temp 250
DEGREES C
Inlet Type PACKED
(OR PURGED
PACKED).
Temp 200
DEGREES C
Operating Mode N/A
Purge Time On N/A min
Purge Time Off N/A
min
Oven Temp Programmed (1 ramp)
Init Temp 110
DEGREES C
Init Time 0
min
Ramp
Rate 15
Fin Temp 150
Fin Time 1
Range 8
COLUMN:
Part No. 19095(#100)
Dimensions 530 MID; 5 M
Sta Phase Methyl
Silicone
FLOW RATES
Carrier (He) 20 +/• 1 ml/min
Hydrogen 33 +/• 1 ml/min
Air 400 +/• 20 ml/min
Makeup (N2) 10 +/• 1 ml/min
Split Vent N/A ml/min
Septum Purge 1•2 ml/min
SAMPLE:
Type FID Sample
Inj Volume 1
l
Part No. 18710•60170
Composition 0.03%(V/V)each
C14,C15,C16 normal paraffin
hydrocarbons in hexane
START
2
1.41
2.55
1.95
C14
C15
C16
HP 5890 Test Sample Operating
Conditions
Flame Ionization Detector (FID)
Figure 10-1.