A SERVICE OF

logo

Test Sample Chromatograms
Test sample chromatograms
232
Detector Type ECDw/MUG
Temp 300 DEGREES C
Inlet Type Split
only or
splitless
Temp 200
DEGREES C
Operating Mode Split(Purge on)
Purge Time On 0 min
Purge Time Off 0
min
Oven Temp Isothermal
Init Temp 170
DEGREES C
Init Time N/A
min
Ramp
Rate 0
Fin Temp
Fin Time
Range 0
COLUMN:
Part No. 19095S(#100)
Dimensions 530 MID; 5 m
Sta Phase Methyl
Silicone
.43
.
7
1.13
2.64
ST
FLOW RATES
Carrier (He) 15 +/• 1 ml/min
Hydrogen N/A ml/min
Air N/A ml/min
Makeup (N2) 60 +/• 2 ml/min
Split Vent 60 +/• 2 ml/min
Septum Purge 5 +/• 1 ml/min
SAMPLE:
Type ECD Sample
Inj Volume 1
l
Part No. 18713•60040
Composition 33 pg/
l (0.033
ppm (W/V)) each: lindane
and aldrin in isooctane
LINDANE
ALDRIN
HP 5890 Test Sample Operating
Conditions
ECD-Split Mode Capillary Inlet
Figure 10-8.