Adjustment Procedures
WFM 90 and WFM 91 Service Manual
5–5
Perform the following steps to adjust the sweep timing and horizontal mag
registration:
1. Replace the signal on the VIDEO IN connector with the output from the
function generator. Be sure to use a 50 W-to-75 W minimum-loss attenuator
and a precision (1%) 50 W cable if the generator specifies a 50 W output.
2. Set the function generator for a 200 kHz sine wave output.
NOTE. Verify the generator frequencies used in these steps by checking them on
the frequency counter.
3. Select 1H SWEEP from the Operating menu.
4. Select 1H SWEEP from the Calibration menu, and then select CAL.
5. Adjust the Arrow Buttons for one cycle per major division.
6. Select CAL MENU, and then select SAVE.
7. Set the function generator frequency to 100 kHz.
8. Select 2H SWEEP from the Operating menu.
9. Select 2H SWEEP from the Calibration menu, and then select CAL.
10. Adjust the Arrow Buttons for one cycle per major division.
11. Select CAL MENU, and then select SAVE.
12. Replace the signal on the VIDEO IN connector with the output from the
leveled sine wave generator. Use the 50 W-to-75 W minimum-loss attenuator
and precision (1%) 50 W cable if the generator specifies a 50 W output.
13. Set the sine wave generator for a 1 MHz output.
14. Select MAG ON from the Operating menu.
15. Select MAG GAIN from the Calibration menu, and then select CAL.
16. Adjust the Arrow Buttons for one cycle per major division.
17. Select CAL MENU, and then select SAVE.
18. Select MAG OFF from the Operating menu.
19. Replace the signal on the VIDEO IN connector with a 75% color bar signal
from the television signal generator, using a 75 W in-line terminator.
20. Select 2F SWEEP from the Operating menu.
21. Select 2F SWEEP from the Calibration menu, and then select CAL.
Sweep Timing and