InstaCal Error Messages
If an InstaCal module that has not been characterized in more than one year is connected to
the Site Master, the follow message will be displayed:
The InstaCal Module Connected, S/N XXXXX
Was Characterized more than 1 year ago.
Characterization date
Current SiteMaster Date
You are advised to return this InstaCal module
to the factory for re-characterization.
Press ENTER to continue, see manual for details.
Return the InstaCal module to the factory for re-characterization in order to ensure contin
-
ued accurate measurements.
If the serial number of the connected InstaCal module does not match the serial number
stored in the Site Master, the following message is displayed:
The InstaCal characterization data stored in the SiteMaster
is for a module different than the one currently connected.
SiteMaster contains data for InstaCal module S/N:
Currently connected InstaCal Module S/N:
Would you like to overwrite the previously
loaded InstaCal characterization?
Press the YES soft key to update the stored InstaCal characterization to use the currently
connected module.
Press the
NO soft key to keep the stored InstaCal characterization.
2-22
Chapter 2 Functions and Operations
Error Message
Description
CANNOT ZERO
NO DETECTOR
INSTALLED
Attempting to perform a Power Monitor zero adjust function with no
RF detector connected to the Site Master.
CANNOT ZERO
INPUT SIGNAL
TOO HIGH
Attempting to perform a Power Monitor zero adjust function with an
input of greater than –20 dBm.
POWER MONITOR
OPTION NOT
INSTALLED
Attempting to enter Power Monitor mode with no Option 5 installed.
Table 2-3. Range Error Messages (2 of 2)