Omega Vehicle Security 1400 CD Player User Manual


 
API Commands Appendix A
A-12 ChartScan User’s Manual
@
- Trigger On Command
TYPE
Acquisition
EXECUTION
Deferred
SYNTAX
@
Trigger upon receipt of an
@
DESCRIPTION
The Trigger On (
@
) command will start or stop the acquisition accordingly when the
start
or
stop
arguments
of the Set Trigger Configuration (
T
) command is set to
1
(trigger on). The
T
command is used as a trigger
source and follows the same rules as other trigger sources as defined by the
T
command. The
@
command is
valid only when an acquisition is configured where either the start and/or stop event is defined as the Trigger On
command. If this command is issued when the acquisition is not configured for stop and/or start Trigger On
command, an error condition will occur.
EXAMPLE
PRINT#1, “OUTPUT07;T1,1,0,0X”
‘ Configure an acquisition with both start and stop
trigger being Trigger On command
PRINT#1, “OUTPUT07; @X”
‘ Start the acquisition
SLEEP 10
‘ Wait 10 seconds to collect data
PRINT#1, “OUTPUT07; @X”
‘ Stop the acquisition