Fluke 45 Turntable User Manual


 
45
Users Manual
6-6
Table 6-2. Performance Tests for Volts, Diode Test, Ohms, and Frequency, Functions (cont.)
Display
Function Range (Rate)
Input Level
Frequency
Min Max
Using decades of 3:
300
3 k
30 k
300 k
3 M
30 M*
300 M*
Short
300
short
3 k
30 k
300 k
3 M
30 M
300 M
Using decades of 1.9:
300
3 k
30 k
300 k
3 M
30 M*
300 M*
Short
190
short
1.9 k
19 k
190 k
1.9 M
19 M
190 M
Using decades of 1:
300
3 k
3 k
30 k
300 k
3 M
30 M*
300 M*
Short
100
short
1 k
10 k
100 k
1 M
10 M
100 M
0.00
299.83
0.0000
2.9983
29.983
299.83
2.9980
29.922
294.0
0.00
189 88
0.0000
1.8988
18.988
189.88
1.8987
18.949
186.2
0.00
99.93
0.0000
0.9993
9.993
99.93
0.9992
9.972
98.0
0.04
300.19
0.0002
3.0017
30.017
300.17
3.0020
30.078
306.0
0.04
190.14
0.0002
1.9012
19.012
190.12
1.9013
19.051
193.8
0.04
100.09
0.0002
1.0007
10.007
100.07
1.0008
10.028
102.0
FREQ 10/100 kHz 0.1 V to 10 V 10 kHz 9.9949 10.006
* Optional test points that can be used if standards are available.