Fluke 45 Turntable User Manual


 
Operating the Meter Using the Computer Interface
Computer Interface Command Set
5
5-25
Range and Measurement Rate Commands and Queries
The commands in Table 5-13 relate to ranging and measurement rates (i.e., readings/
second). In the autorange mode, the meter automatically selects a range for each reading;
in the manual range mode, the user selects a fixed range.
Table 5-13. Range and Measurement Rate Commands and Queries
Command Description
Causes the meter to enter the autoranging mode on the primary display. If the
autorange mode cannot be selected (e.g., if REL, dB, MN MX, or diode/continuity test
is selected), an Execution Error is generated.
Causes meter to return "1" if it is in autorange, or "0" if it is not.
Causes the meter to exit autoranging on the primary display and enter manual ranging.
The present range becomes the selected range.
Sets the primary display to the range designated by <range>. <range> is a number
between 1 and 7 that corresponds to a range shown in Tables 5-12A and 5-12B.
Table 5-13A. Ranges at Fast & Medium Measurement Rate
Range
Value
Voltage
Range
Ohms
Range
Current
Range
Frequency
Range
1
2
3
4
5
6
7
300 mV
3 V
30 V
300 V
1000 V dc*
ERROR
ERROR
300
3 k
30 k
300 k
3 M
30 M
300 M
30 mA
100 mA
10 A
ERROR
ERROR
ERROR
ERROR
1000 Hz
10 kHz
100 kHz
1000 kHz
1 MHz
ERROR
ERROR
* 1000 V dc, 750 V ac
Table 5-13B. Ranges at Slow Measurement Rate
Range
Value
Voltage
Range
Ohms
Range
Current
Range
Frequency
Range
1
2
3
4
5
6
7
100 mV
1000 mV
10 V
100 V
1000 V dc*
ERROR
ERROR
100
1000
10 k
100 k
1000 k
10 M
100 M
10 mA
100 mA
10 A
ERROR
ERROR
ERROR
ERROR
1000 Hz
10 kHz
100 kHz
1000 kHz
1 MHz
ERROR
ERROR
* 1000 V dc, 750 V ac
AUTO
AUTO?
FIXED
RANGE
<range>