Fluke 45 Turntable User Manual


 
Appendices
Specifications A
A-7
Ohms
Resolution
Range
Slow Medium Fast
Accuracy
Typical Full
Scale
Voltage
Max Current
Through the
Unknown
300
3 k
30 k
300 k
3 M
30 M
300 M*
100
1000
10 k
100 k
1000 k
10 M
100 M*
1 m
10 m
100 m
1
10
100
100 k
10 m
100 M
1
10
100
1 k
100 k
100 M
1
10
100
1 k
10 k
1 M
0.05 % + 2 + 0.02
0.05 % + 2
0.05 % + 2
0.05 % + 2
0.06 % + 2
0.25 % + 3
2 %
0.05 % + 8 + 0.02
0.05 % + 8 + 0.02
0.05 %+8
0.05 % + 8
0.06 % + 8
0.25 % + 6
2 % + 2
0.25
0.24
0.29
0.29
0.3
2.25
2.9
0.09
0.10
0.11
0.11
0.12
1.5
2.75
1 mA
120µA
14 µA
1.5 µA
150 µA
320 µA
320 µA
1 mA
120 µA
14 µA
1.5 µA
150 µA
150 µA
320 µA
*Because of the method used to measure resistance, the 100 M (slow) and 300 M (medium and fast)
ranges cannot measure below 3.2 M and 20 M, respectively. "UL" (underload) is shown on the display for
resistances below these nominal points, and the computer interface outputs "+1 E-9".
Open Circuit Voltage
3.2 V maximum on the 100 , 300 , 30 M, 100 M, and 300 M ranges, 1.5 V
maximum on all other ranges.
Input Protection
500 V dc or rms ac on all ranges
Diode Test/Continuity
Maximum Reading Resolution
Slow
Medium
Fast
999.99 mV
2.5 V
2.5 V
10 µV
100 µV
1 mV
Test Current
Approximately 0.7 mA when measuring a forward biased junction.
Audible Tone
Continuous tone for continuity. Brief tone for normal forward biased diode or
semiconductor junction.