Measurement Cable Connections
Voltage Sensing at a Location Other than the External Current Shunt
You may also need to open up the cable connector if you need to voltage sense at a point other than at the
external current shunt. You must use your own sense cable to make the voltage sense connections. Please
completely remove the voltage sense wires from the Agilent supplied measurement cable if you are using
your own remote voltage sense cable. Disconnect the voltage sense wires from both cable connectors.
To use your own cable to voltage sense at a point other than the external current shunt, make the
connections at the point where you will be remote sensing. Insert the other end of your cable into the HI
and LO voltage sense terminals of the ac source connector. The cable connectors accepts wire sizes from
AWG 22 to AWG 12. Firmly tighten the screws when making wire connections.
WARNING: SAFETY HAZARD You must replace the cable hoods after completing all connections.
This is because the connector screw terminals will be at line potential during operation.