Agilent Technologies 5964-8134 Turntable User Manual


 
23
Verification Procedure
The following verification procedures assume you understand how to operate the ac source from the front
panel as explained in chapter 4.
Perform the following tests for operation verification in the order indicated.
1. Turn-On Checkout
2. Voltage Measurement Accuracy
3. Current Measurement Accuracy
NOTE: The ac source must pass turn-on selftest before you can proceed with the verification.
Test Set up
COM
R
Rs
L
Rs = Current Monitor resistor (0.01 ohms)
Agilent 3458A
DMM
COM
01
SENSE
R = Load resistor for CC test (30 ohms)
01
Agilent 3458A
DMM
Agilent 6811B/6812B/6813B
MAIN OUTPUT
POWER ANALYZER
LO HI | LO HI
I
SENSE
V
SENSE
Agilent 33120A
FUNCTION GENERATOR
COM
COM
01
SENSE
01
Agilent 6811B/6812B/6813B
MAIN OUTPUT
POWER ANALYZER
LO HI | LO HI
I
SENSE
V
SENSE
Agilent 3458A
DMM
C1
C1 = Capacitor (1.0 uF)
A.
B.
LO
HI
LO
HI
LO
HI
L
Figure B-1. Verification / Calibration Test Setup