14
Additional Calibration Commands
The following SCPI commands are used to calibrate the Power Analyzer input and external current
shunt:
CALibrate:VOLTage:EXTernal
Initiates calibration of the external voltage measurement.
Command Syntax
CALibrate:VOLTage:EXTernal
Parameters
None
Examples
CAL:VOLT:EXT
CALibrate:CURRent:EXTernal
Initiates calibration of the external current measurement.
Command Syntax
CALibrate:CURRent:EXTernal
Parameters
None
Examples
CAL:CURR:EXT
CALibrate:SHUNt
Lets you enter the external current shunt value (in ohms). The external current shunt is used for external
current measurement. The programming range is 1E-6 to 1E6. This parameter is nonvolatile, so there is
no *RST value. As with other CAL commands, calibration must first be enabled with CAL:STATe, and
the value must be made permanent with CAL:SAVE.
Command Syntax
CALibrate:SHUNt <value>
Parameters
Current shunt value in ohms
Examples
CAL:SHUN .01