IMPORTANT NOTICE -- ERRATA
Model 1154 Product Manual 00809-0100-4514 Rev BA (January 2008)
No.
Affected
Pages
Description of Change
Effect.
Date
1 6-6
Process Flange –CF3M (Cast version of 316L SST)
Drain/Vent Valves –316L SST
Process Connections – 3/8-inch Swagelok compression fitting, 316L SST (1/4-18
NPT optional)
10/21/09
2 3-5 Change the last paragraph on the page to read as follows:
”Damping electronics are available as an option. Transmitters with standard
electronics can be retrofitted with the adjustable damping feature by changing out
both the amplifier board and the calibration board. Please reference Table 6-2
(Rosemount 1154DP, 1154HP, and 1154GP Spare Parts) for the applicable part
numbers.”
4/13/12
3 6-9 Table 6-2,”Rosemount 1154DP, 1154HP, and 1154GP Spare Parts”, in all locations
the following part numbers are updated:
Amplifier Circuit Board, Output Code R:
”01154-0001-0005” is replaced by ”01154-0153-0001”
Amplifier Circuit Board with Damping, Output Code R:
”01154-0021-0004” is replaced by ”01154-0156-0001”
Amplifier Circuit Board, Output Code R, N0026:
”01154-0001-0006” is replace by ”01154-0153-0002”
Sensor Module, 316 SST
(5)
:
0-25/150 inH
2
O: ”01154-0300-0242” is replaced by ”01154-5300-0242”
”01154-0300-0342” is replaced by ”01154-5300-0342”
”01154-0300-0142” is replaced by ”01154-5300-0142”
0-125/750 inH
2
O: ”01154-0300-0252” is replaced by ”01154-5300-0252”
”01154-0300-0352” is replaced by ”01154-5300-0352”
”01154-0300-0152” is replaced by ”01154-5300-0152”
0-17/100 psi: ”01154-0300-0262” is replaced by ”01154-5300-0262”
”01154-0300-0362” is replaced by ”01154-5300-0362”
”01154-0300-0162” is replaced by ”01154-5300-0162”
0-50/300 psi: ”01154-0300-0272” is replaced by ”01154-5300-0272”
”01154-0300-0372” is replaced by ”01154-5300-0372”
”01154-0300-0172” is replaced by ”01154-5300-0172”
0-170/1,000 psi: ”01154-0300-0282” is replaced by ”01154-5300-0282”
”01154-0300-0182” is replaced by ”01154-5300-0182”
0-500/3,000 psi: ”01154-0300-0192” is replaced by ”01154-5300-0192”
0-1,000/4,000 psi: ”01154-0300-0102” is replaced by ”01154-5300-0102”
4/13/12
4 6-10 Table 6-2,” Rosemount 1154DP, 1154HP, and 1154GP Spare Parts”, the table
footnotes are updated to add note (5) which will read as follows:
(5) IMPORTANT NOTICE: To maintain a transmitter’s qualified configuration,
when purchasing or installing a new Sensor Module, Rosemount Qualification
report D2011019 must be carefully reviewed to verify that the Sensor Module
to be installed and the associated Amplifier Circuit Board in a given
transmitter is a qualified configuration. As detailed in the referenced report, not
all Sensor Module part numbers are considered qualified in conjunction with certain
Amplifier Circuit Boards.
4/13/12