Incoming Inspection
1-22
CSA7000 Series, TDS7000 Series, & TDS6000 Series Instruments User Manual
H One TCA-BNC TekConnect adapter , or one TCA-SMA TekConnect adapter
and one SMA male-to-BNC female adapter , such as Tektronix part number
015-1018-xx
Self Tests
This procedure uses internal routines to verify that the instrument functions and
was adjusted properly. No test equipment or hookups are required.
Equipment
required
None
Prerequisites Power on the instrument and allow a 20 minute warm-up before doing
this procedure.
1. Verify that internal diagnostics pass: Do the following substeps to verify
passing of internal diagnostics.
a. Display the System diagnostics menu:
H If the instrument is in tool-bar mode, click the MENU button to put
the instrument into menu bar mode.
H From the Utilities menu, select Instrument Diagnostics . . . . This
displays the diagnostics control window.
b. Run the System Diagnostics:
H First disconnect any input signals and probes from all four channels.
H Click the Run button in the diagnostics control window.
c. Wait: The internal diagnostics do an exhaustive verification of proper
instrument function. This verification will take five to fifteen minutes.
When the verification is finished, the resulting status will appear in the
diagnostics control window.
d. Verify that no failures are found and reported on-screen. All tests should
pass.