Fluke 7000DP CD Player User Manual


 
Impulse 6000D/7000DP Users Manual Supplement
4/08 3
Scope Output
Autorange ...........................................................2000:1, 400:1 and 80:1: dependant on the range
Waveform Playback
Output.............................................................BNC
Output impedance...........................................50 Ω
Amplitude Accuracy........................................±5 %
Charge Time Measurement
Range .................................................................0.1 to 100.0 s
Accuracy.............................................................±0.05 s, typical
Synchronization Test (Elective Cardioversion)
Delay Time Measurement
Timing window................................................ECG R-wave peak to the defib pulse peak
Range .............................................................-120 to +380 ms; measures timing from 120 ms prior to the R-wave peak to up to 380
ms following the R-wave peak.
Resolution.......................................................1 ms
Accuracy.........................................................±1 ms
ECG waves
Normal Sinus Rhythm (NSR)..........................10 to 180 (by 1) BPM
Atrial fibrillation................................................Coarse and fine
Monomorphic Ventricular Tachycardia ...........120 to 240 (by 5) BPM
Asystole ..........................................................Flat line
Automated Defibrillator Test ECG Waves
Normal Sinus ......................................................10 to 300 (by 1) BPM