Fluke 7000DP CD Player User Manual


 
Manual Supplement Impulse 6000D/7000DP Users
2 8/08
Accuracy.............................................................±0.1 ms
Voltage
Range.................................................................20 to 5000 V
Accuracy.............................................................±(1 % of reading + 2 V)
Current
Range.................................................................0.4 to 100.0 A
Accuracy.............................................................±(1 % of reading + 0.1 A)
Tilt (biphasic and pulsed biphasic)
Range................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Interphase delay (biphasic and pulsed biphasic)
Range................................................................. 0.1 ms to 9.9 ms
Accuracy............................................................. ±0.1 ms
Frequency (pulsed biphasic only)
Range................................................................. 2000 Hz to 8000 Hz
Accuracy............................................................. ±1 % of reading
Duty cycle (pulse biphasic only)
Range................................................................. 1 % to 99 %
Accuracy............................................................. ±1 digit
Sample rate ...........................................................250 kHz (4 μs sample)
Maximum Average Power....................................12 W, equivalent to 10 defib pulses of 360 J every 5 minutes