Impulse 6000D/7000DP Users Manual Supplement
4/08 1
Change #1, 46470, 46697, 47714, 48007, 49680
Replace pages 31 through 35, with the following:
Defibrillator Analyzer Specifications
Energy Output Measurement
Compatible Defibrillator Waveshapes ................Lown, Edmark, Trapezoidal, DC Bi-phasic, and AC Pulsed Bi-phasic
Note
AC Pulsed Biphasic waveform has not been approved in the United States.
Autoranged Measurement....................................0.1 to 600 J
Accuracy
0.1 to 360 J.....................................................±(1 % of reading + 0.1 J)
360 to 600 J....................................................±(1 % of reading + 0.1 J), typical
Note
For Pulsed Bi-Phasic defibrillator, specified accuracy is
±
(1.5 % of reading + 0.3 J) on both ranges.
Load resistance
Resistance..........................................................50 Ω
Accuracy.........................................................±1 %, non-inductive (<2 μH)
Pulse trigger level.................................................20 V
Pulse width
Range .................................................................1.0 to 50.0 ms