Fluke 7000DP CD Player User Manual


 
Impulse 6000D/7000DP Users Manual Supplement
4/08 9
Pulse
playback
+4 V, 1 ms pulse Start of defibrillator pulse playback
[1]
[1] Pulse playback includes 0.4 ms before the pulse was detected and therefore displays any early
transitional signals.
Use the synchronizing marker pulses to display specific information from the Scope Output signal. To see when the
defibrillator fires, set the oscilloscope trigger level to +2 V. To get the ranging information, set the oscilloscope to trigger
on the -4 V markers. To ignore the ranging information and see only the defibrillator pulse, set the oscilloscope to trigger
on levels above +2 volts.
On page 26, prior to Controlling the Analyzer Remotely, add the following section:
Setting the ECG Reference Lead
ECG signal amplitude is set by designating either lead I or lead II as the reference lead. Lead II is the default
selection. To change the reference lead, do the following:
Press Θ to reveal the main setup menu. Next press the softkey labeled More for additional setup
selections. Press the softkey labeled ECG Ref Lead to toggle between lead I and lead II. Once the
reference lead is set, press the softkey labeled Back or one of the other functions to store the selection
in non-volatile memory.